We provide platelet shakers (registered medical device: blood refrigerator) for Japanese Red Cross Society’s blood centers and large-scale hospitals in Japan. Our shaking technology, which we have accumulated over many years in pharmaceutical and food industries, boasts robustness and stability that is unrivaled by others. A unique drive system that slides the stand on four shafts enables low noise and high durability. UPS unit is included as standard, and in the event of a power outage, it will issue a power outage alarm and record the temperature for a certain period of time.
Features
- Easy operation with touch panel
- Trays that prevents platelet bags from falling out
- Warning function for door opening and shaking stop
- Accurate temp. controls
- Two completely independent temp. monitor systems
Specification
Model | PRK20 |
Shaking mode | Reciprocal |
Shaking speed | 50 – 120rpm(Common: 60±3prm) |
Shaking width | 50mm |
Tray stages | 20 |
Bag capacity | 8 bag per tray x 20 stages = 160 bags * In the case of 1000ml PC (160×270mm) |
Temp. range | 15.0 – 40.0℃(Common: 22.0℃) *Ambient range: 10 – 30℃ |
Temp. accuracy (measurement) | ±0.5℃ |
Temp. accuracy (distribution) | 22.0±0.1℃ |
Safety systems | Short circuit breaker, Open door/lid cut-off switch, Overtemperature protection, Overheat protection in refrigerator |
Alarms | Control panel alarm display (shaker abnormality, overtemperature, fan abnormality, refrigerator abnormality, overload abnormality, door opening abnormality) and alarm buzzers |
Power outage alarm | UPS |
Tray mateiral | SUS304 |
Power | AC100V 50/60Hz |
Weight | Approx. 700kg |
External dimensions | W1420×D805×H1797 |